Technical Specifications

Isolation Dynamic: 1.2Hz - 200Hz
purely passive beyond 200Hz
Transmissibility Comparison transmissibility curve AVI series vs. passive systems
Transmissibility above 10 Hz is always less than 0,017 (-35dB)
  Transmissibility curve AVI series horizontalTransmissibility curve AVI series vertical
Correction Forces +/- 16 N vertical
+/-  8 N horizontal
Static Compliance approx. 1.75 µm/N vertical
approx. 3.5  µm/N horizontal
Maximum load 800 kg (single unit: 400 kg)
Dimensions (single unit): Top plate :
360 x 190 mm (AVI-400S)
720 x 190 mm (AVI-400M)
820 x 230 mm (AVI-400XL)

Bottom plate :
396 x 182 mm (AVI-400S)
636 x 182 mm (AVI-400M)
856 x 182 mm (AVI-400XL)

Height :
112.5 mm
Weight (single unit): 20 kg (AVI-400S)
24 kg (AVI-400M)
28 kg (AVI-400XL)

A4CU Controller

Safety class 1
Power consumption Typically 18 W 
Input voltage 90-125VAC / 200-250VAC +/- 10% , 50 - 60 Hz
Overload 16 LEDs indicate overload in input stages
Monitor signal A multiplexed signal for display on oscilloscopes shows vibration levels with and without isolation - for diagnostic purposes only

Normal Environmental Conditions

Protection class IP 20
Temperature range 5°C - 40°C
Relative humidity 10-90 % (5-30 °C) / 10-60 % (30-40 °C)
Application Indoor
Altitude Up to 2000 m (6.500 ft)